MEASure? and CONFigure Defaults
Setting SCPI Command System Defaults Used
Common
to all
Functions
Input Channel ROUTe:TERMinals Channel 1
Digital Filter INPut:FILTer Off
Digital Filter Precharge INPut:FILTer Last value set
Analog Filter INPut:FILTer Off
Trigger Source TRIGger:SOURce Immediate
Trigger Delay TRIGger:DELay Auto
Trigger Count TRIGger:COUNt 1
N Samples SAMPle:COUNt 1
Math CALCulate Off
Readings Store DATA:FEED RDG_STORE "CALC"
Display (On or Off) DISPlay On
Specific
To Each
Function
Integration Time
(NPLC)
<
resolution
>(10 plc)
Range <
range
> Autorange
Null SENSe:
<function>
:NULL Off
Resistance
Specific
Offset Compensated
SOURce:FRESistance:OCOMPensated Off
Voltage Limited
SOURce:FRESistance:VOLTage:LIMit Last value set
Low Power
SOURce:FRESistance:POWer:LIMit Last value set
Temperature
Specific
Transducer SENSe:TEMPerature:TRANsducer Last probe set
Probe Type SENSe:TEMPerature:TRANsducer Last type set
RTD type and resistance SENSe:TEMPerature:TRANsducer Last type set
The MEASure? and CONFigure commands use default parameters for
volatile memory parameters. Non-volatile memory parameters are not
modified and the MEASure? or CONFigure commands use parameters
in memory. See page 112. For example, the temperature units last set
are stored in non-volatile memory
and are not changed by a MEASure? or CONFigure command.
4
Chapter 4 Remote Interface Reference
Simplified Programming Sequence
127