Automatic Trigger Delays
If you do not specify a trigger delay, the meter selects an automatic delay
for you. The delay is determined by function, range, and integration time.
DC voltage :
Range
NPLC
<
1
Trigger Delay
NPLC
³
1
Trigger Delay
1
m
V
10
m
V to 120 V
15 ms
1.0 ms
15 ms
1.5 ms
Resistance (2-wire and 4-wire):
Range NPLC < 1
Trigger Delay
NPLC
³
1
Trigger Delay
1
10
100
1 k
10 k
100 k
1 M
1.0 ms
1.0 ms
1.0 ms
1.0 ms
1.0 ms
4 ms
40 ms
1.5 ms
1.5 ms
1.5 ms
1.5 ms
1.5 ms
6 ms
60 ms
Temperature:
NPLC < 1
Trigger Delay
NPLC
³
1
Trigger Delay
RTD
THERmistor
ThermoCouple
1.0 ms
1.0 ms
1.0 ms
1.5 ms
1.5 ms
1.5 ms
3
Chapter 3 Features and Functions
Triggering
89