Using the range and resolution Parameters
With the MEASure? and CONFigure commands, you can select the
measurement function, range, and resolution all in one command. Use
the range parameter to specify a fixed range larger than the expected
value of the input signal. You can set the range parameter to AUTO to
set autoranging.
Use the resolution parameter to specify the desired resolution for the
measurement. Setting the resolution sets the integration time of the
meter. The resolution parameter can be set to any number but the
resulting integration time (NPLC) is one of the following:
<
resolution
>
Parameter
Integration Time
(NPLC)
< 0.0001 x Range 0.02
< 0.00001 x Range .2
< 0.000003 x Range 1
< 0.0000022 x Range 2
< 0.000001 x Range 10
< 0.0000008 x Range 20
< 0.0000003 x Range 100
< 0.00000022 x Range 200
Specify the resolution in the same units as the measurement function,
not in number of digits. For example, for volts measurements, specify the
resolution in volts. For resistance, specify the resolution in ohms.
You must specify a range to use the resolution parameter.
4
Chapter 4 Remote Interface Reference
Simplified Programming Sequence
129

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34420A and is the answer not in the manual?

Agilent Technologies 34420A Specifications

General IconGeneral
BrandAgilent Technologies
Model34420A
CategoryMultimeter
LanguageEnglish

Related product manuals