To Set the Parity
You can select the parity for RS-232 operation. The meter is configured for
even parity with 7 data bits when shipped from the factory. See also
“Parity Selection,” on page 104.
On/Off
Shift < 1 Turn on the front-panel menu.
A: MEAS MENU
< < 2 Move across to the I/O MENU choice on this level.
F: I/O MENU
< < 3 Move down a level and then across to the PARITY command.
4: PARITY
4 Move down to the “parameter” level to select the parity.
Use the left/right arrow keys to see the parity choices. Choose from one of
the following: None (8 data bits), Even (7 data bits), or Odd (7 data bits).
When you set parity, you are indirectly setting the number of data bits.
EVEN: 7 BITS
Auto/Man 5 Save the change and turn off the menu.
ENTER The parity selection is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.
Remote Interface: you can set or query the parity using the following
commands:
SYStem:COMMunicate:SERial
:PARity {EVEN | ODD | NONE}
:PARity?
4
Chapter 4 Remote Interface Reference
To Set the Parity
191

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34420A and is the answer not in the manual?

Agilent Technologies 34420A Specifications

General IconGeneral
BrandAgilent Technologies
Model34420A
CategoryMultimeter
LanguageEnglish

Related product manuals