RS-232 Interface Commands
SYSTem:LOCal
This command places the meter in the local mode during RS-232
operation. All keys on the front panel are fully functional.
SYSTem:REMote
This command places the meter in the remote mode during RS-232
operation. All keys on the front panel except the LOCAL key are
disabled. You MUST send this command before any other RS-232
command can be received.
SYSTem:RWLock
This command places the meter in the remote mode during RS-232
operation. All keys on the front panel are disabled, including the LOCAL
key.
Caution
Sending or receiving data over the RS-232 interface when not configured
for REMOTE operation can cause unpredictable results. Always make
sure the meter is configured for remote operation when using the RS-232
interface.