Loading...
[SENSe1: | SENSe2:]VOLTage[:DC]:NPLCycles
{0.02 | 0.2 | 1 | 2 | 10 | 20 | 100 | 200 | MIN | MAX}
[SENSe:]
FRESistance | RESistance
:NPLCycles {0.02 | 0.2 | 1 | 2 | 10 | 20 | 100 | 200 | MIN | MAX}
[SENSe:]TEMPerature:NPLCycles {0.02 | 0.2 | 1 | 2 | 10 | 20 | 100 | 200 | MIN | MAX}
This command selects the integration time in number of power line cycles
for the present function (the default is 10 PLC). MIN = 0.02. MAX = 200.
[volatile memory]
[SENSe1: | SENSe2:]VOLTage[:DC]:NPLCycles? {MIN | MAX}
[SENSe:]
FRESistance | RESistance
:NPLCycles? {MIN | MAX}
[SENSe:]TEMPerature:NPLCycles? {MIN | MAX}
This command queries the integration time for the selected function.
4
Chapter 4 Remote Interface Reference
Setting the Function, Range, and Resolution
139