[SENSe:]TEMPerature:TRANsducer:TYPE {TC | THER | FRTD | DEF}
This command sets the type of temperature transducer to use for
temperature measurements. Choose TC (thermocouples), THER
(thermistors), or FRTD (four-wire RTD). DEF sets FRTD.
[SENSe:]TEMPerature:TRANsducer:TYPE?
This command queries for the current temperature measurement
transducer type.
[SENSe:]TEMPerature:TRANsducer:TCouple:TYPE {B | E | J | K | R | S | T | DEF}
This command sets the type of thermocouple to use for the temperature
measurement. DEF sets the last thermocouple type used.
[SENSe:]TEMPerature:TRANsducer:TCouple:TYPE?
This command queries for the thermocouple type in use.
[SENSe:]TEMPerature:TRANsducer:TCouple:RJUNction {<
reference
> | MIN | MAX}
This command sets the reference junction temperature for thermocouple
measurements when using a fixed temperature reference.
MIN sets 0° C, MAX sets 55° C.
[SENSe:]TEMPerature:TRANsducer:TCouple:RJUNction? [{MIN | MAX}]
This command queries for the reference junction temperature, if set,
when using a fixed temperature reference.
MIN returns 0° C, MAX returns 55° C.
[SENSe:]TEMPerature:TRANsducer:TCouple:RJUNction:TYPE {FIX | INT | THER}
This command sets the reference junction type for thermocouple
measurements. The type can be se to a fixed value (FIX), can be set to use
the internal thermistor (INT), or can use a thermistor connected to
channel 1 (THER). When type is set to FIX, use
SENSe:TEMPerature:TRANsducer:TCouple:RJUNction <
value
>
to set the reference temperature.
[SENSe:]TEMPerature:TRANsducer:TCouple:RJUNction:TYPE ?
This command queries for the type of reference junction being used for
thermocouple measurements.