EasyManua.ls Logo

Agilent Technologies 34420A - Page 160

Agilent Technologies 34420A
294 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SYSTem:ERRor?
This command queries the meter’s error queue. Up to 20 errors can be
stored in the queue. Errors are retrieved in first-in-first out (
FIFO) order.
Each error string may contain up to 80 characters.
SYSTem:VERSion?
This command queries the meter to determine the present SCPI version.
Returns "1994.0"
*RST
This command resets the meter to its power-on configuration.
*TST?
This command performs a complete self-test of the meter. Returns “0” if
the self-test is successful, or “1” if the test fails.
*IDN?
This command reads the meter’s identification string (be sure to
dimension a string variable with at least 40 characters).
Chapter 4 Remote Interface Reference
System-Related Commands
158

Table of Contents

Related product manuals