The questionable data event register is cleared when:
You execute a *CLS (clear status) command.
You query the event register using STATus:QUEStionable:EVENt?.
The questionable data enable register is cleared when:
You turn on the power (*PSC does not apply).
You execute the STATus:PRESet command.
You execute the STATus:QUEStionable:ENABle 0 command.