*PSC {0 | 1}
(Power-on status clear.) This command clears the Status Byte and
Standard Event register enable masks when power is turned on (*PSC
1). When *PSC 0 is in effect, the Status Byte and Standard Event
register enable masks are not cleared when power is turned on.
[non-volatile memory]
*PSC?
This command queries the power-on status clear setting.
Returns “0” (*PSC 0) or “1” (*PSC 1).
*SRE <enable value>
This command enables bits in the Status Byte enable register.
*SRE?
This command queries the Status Byte enable register. The meter
returns a decimal value which corresponds to the binary-weighted sum of
all bits set in the register.
4
Chapter 4 Remote Interface Reference
Status Reporting Commands
173