EasyManuals Logo
Home>Agilent Technologies>Multimeter>34420A

Agilent Technologies 34420A User Manual

Agilent Technologies 34420A
294 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #278 background imageLoading...
Page #278 background image
Triggering and Memory
Reading HOLD Sensitivity 10%, 1%, 0.1%, or 0.01% of range
Samples/Trigger 1 to 50,000
Trigger Delay
0 to 3600s; 10µs step size
External Trigger Delay <1ms
External Trigger Jitter
<500µs
Memory 1024 readings
Math Functions
NULL (Channel 1 DCV, Channel 2 DCV, Difference,
MMMNResistance, Temperature)
STATS (Min, Max, Average, Peak-Peak, Standard Deviation,
MMMMIINumber of readings)
SCALE (Allows linear scaling as y=mx+b)
CHART NULL (Establishes zero for rear panel output)
Filter (Analog or Digital or Both)
Analog Low pass 2 pole @ 11Hz, available
for 1, 10, 100mV range
Digital Moving average filter, 10 (fast),
50 (medium) or, 100 (slow) reading
averages.
Chart Out (Analog Out)
Maximum output: ± 3V
Accuracy: ± 0.1% of output + 1 mV
Output Resistance: 1 kΩ ± 5%
Update rate: once per reading
Span and Offset: Adjustable
Standard Programming Languages
SCPI (IEEE 488.2), Keithley 181
Accessories Included
4 ft low thermal cable with copper spade lugs, 4-wire shorting
plug, user’s manual, service manual, contact cleaner, test
report and power cord
General Specifications
Front Panel Connection shielded, low thermal, 99% copper
contacts
Power Supply
100V/120V/220V(230V)/240V ± 10%
Power Line Frequency 45Hz to 66Hz and 360Hz to 440Hz
Automatically sensed at power-on.
Power Consumption 25VA peak (10W average)
Operating Environment
Full accuracy for 0° C to 55° C
Full accuracy to 80% R.H. up to 40° C
Storage Environment
-40° C to 75° C
Weight 3 kg (6.5 lbs)
Safety Designed to CSA, UL-1244,
IEC-1010
RFI and ESD CISPR 11
Vibration and Shock MIL-T-28800E, Type III, Class 5
Warranty 3 year
8
Chapter 8 Specifications
34420A Nano Volt/Micro Ohm Meter
279

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34420A and is the answer not in the manual?

Agilent Technologies 34420A Specifications

General IconGeneral
BrandAgilent Technologies
Model34420A
CategoryMultimeter
LanguageEnglish