• Front Panel operation: Integration time can be set indirectly when
you select the number of digits (See page 43).
You can also set the integration time in the MEASurement menu with
the 2: INTEGRATE command.
• Remote operation: Refer to the table on page 129. Use one of the
following commands:
[SENSe:]
VOLT:DC:NPLC {0.02|0.2|1|2|10|20|100|200|MIN|MAX}
FRES|RES
:NPLC {0.02|0.2|1|2|10|20|100|200|MIN|MAX}
TEMP:NPLC {0.02|0.2|1|2|10|20|100|200|MIN|MAX}
Integration time and resolution are related. The CONFigure and
MEASure? commands can also set the integration time, see page 129.
Integration Time is local to the selected function. This means that you
can select the integration time for each function independently. The
meter remembers the integration time when you switch between
functions.
3
Chapter 3 Features and Functions
General Measurement Configuration
59