Agilent 4155C/4156C Users Guide Vol.1, Edition 11 8-13
Specifications
Hardware
41501B SMU and Pulse Generator Expander
HPSMU (High Power SMU) Specifications
Current Range, Resolution, and Accuracy
See Table 8-6.
Voltage Range, Resolution, and Accuracy
See Table 8-7.
Output terminal/connection
Dial triaxial connectors, Kelvin (remote sensing)
Voltage/Current Compliance (Limiting)
The SMU can limit output voltage or current to prevent damaging the device
under test.
Voltage 0 V to ± 200 V
Current ± 1 pA to ± 1 A
Compliance accuracy Same as current (voltage) set accuracy.
HPSMU supplemental information
Allowable cable resistance (Kelvin connection.)
Force: maximum 0.7 Ω (100 mA to 1A)
Force: maximum 10 Ω (
100 mA)
Sense: maximum 10 Ω
Voltage source output resistance typical 0.2 Ω (Force line. Non Kelvin
connection.)
Voltage measurement input resistance 10
13
Ω
Current source output resistance 10
13
Ω (1 nA range)
Current compliance setting accuracy (for opposite polarity)
For 1 nA to 10 nA range:
V/I setting accuracy ± 12 % of range
For 100 nA to 1 A range:
V/I setting accuracy ± 2.5 % of range

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 41501B and is the answer not in the manual?

Agilent Technologies 41501B Specifications

General IconGeneral
BrandAgilent Technologies
Model41501B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals