8-16 Agilent 4155C/4156C User’s Guide Vol.1, Edition 11
Specifications
Hardware
PGU (Pulse Generator Unit) Specifications
General
Modes pulse or constant
Amplitude 0 Vpp to 40 Vpp
Window -40.0 V to +40.0 V
Maximum current
± 200 mA (width: 1 ms, average current 100 mA)
± 100 mA
Output impedance 50 Ω and low impedance (
1 Ω )
Burst count range 1 to 65535
Trigger output Level: TTL
Timing: same timing and width as PGU1 pulse output.
Pulse/dc Output Voltage and Accuracy
See Table 8-8.
Timing Parameters
See Table 8-9.
Accuracy of Timing parameters
The following data is applied when leading time and trailing time are the same
value.
Period
± (2 % + 2 ns)
Width
± (3 % + 2 ns)
Delay
± (2 % + 40 ns)
Leading Time and Trailing Time
See Table 8-10.
Restrictions of Pulse Timing Parameters Setup
Pulse width < Pulse period
Delay time < Pulse period
Leading time < Pulse width × 0.8
Trailing time < (Pulse period - Pulse width) × 0.8