Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 87
Maintenance 2
To clean the APCI spray chamber daily
1 Turn off the spray chamber.
2 Remove the corona needle.
3 Make sure the needle has cooled and then carefully clean it with abrasive
paper.
4 Open the spray chamber.
Figure 34 Opened spray chamber
5 Rinse the interior of the spray chamber with the current mobile phase or
with a mixture of isopropanol and water.
6 Wipe the interior of the spray chamber and the end of the vaporizer with a
clean cloth.
7 Remove the spray shield.
8 Use abrasive paper to gently clean the end of the capillary cap.
9 Dampen a clean cloth and wipe the end of the capillary cap.
NOTE
Recent residue should be soluble in the mobile phase. If you are not sure what mobile
phase was used recently, use a mixture of 50% isopropanol and 50% water as a general
cleaning solution.
Use the weekly cleaning procedure if symptoms of contamination persist, or if the spray
shield or capillary cap shows significant discoloration that cannot be removed by the
regular daily cleaning.