Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 87
Maintenance 2
To clean the APCI spray chamber daily
1 Turn off the spray chamber. 
2 Remove the corona needle. 
3 Make sure the needle has cooled and then carefully clean it with abrasive 
paper. 
4 Open the spray chamber.
Figure 34 Opened spray chamber
5 Rinse the interior of the spray chamber with the current mobile phase or 
with a mixture of isopropanol and water. 
6 Wipe the interior of the spray chamber and the end of the vaporizer with a 
clean cloth.
7 Remove the spray shield.
8 Use abrasive paper to gently clean the end of the capillary cap.
9 Dampen a clean cloth and wipe the end of the capillary cap.
• Recent residue should be soluble in the mobile phase. If you are not sure what mobile 
phase was used recently, use a mixture of 50% isopropanol and 50% water as a general 
cleaning solution.
• Use the weekly cleaning procedure if symptoms of contamination persist, or if the spray 
shield or capillary cap shows significant discoloration that cannot be removed by the 
regular daily cleaning.