EasyManuals Logo

Agilent Technologies 6500 Series User Manual

Agilent Technologies 6500 Series
154 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #106 background imageLoading...
Page #106 background image
106 Agilent 6500 Series Q-TOF LC/MS Maintenance Guide
2Maintenance
To adjust the multimode nebulizer
To adjust the multimode nebulizer
When required When the needle is plugged. Common symptoms of a plugged needle are
increased LC back pressure or off-axis spraying or dripping from the
nebulizer.
Tools required Adjustment fixture (p/n G1946-20215)
Gloves, clean
Pliers, long nose (p/n 7810-0004)
Wrench 3-mm, open-end (p/n 8710-2699)
Wrench ¼-inch x 5/16-inch, open-end (p/n 8710-0510)
Parts required Nebulizer needle kit, Electrospray (includes needle, ferrule, and needle holder,
p/n G2427A)
1 Install the nebulizer in the adjustment fixture.
2 Loosen the locknut next to the zero-dead-volume (ZDV) union.
3 Remove the union from the nebulizer.
4 Loosen the locknut of the needle holder.
5 Unscrew the needle holder and pull it out of the nebulizer.
6 Slide the non-tapered end of the needle through the new needle holder from
the narrower side.
Figure 47 Electrospray nebulizer needle in needle holder

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6500 Series and is the answer not in the manual?

Agilent Technologies 6500 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6500 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals