Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 151
Reference 3
Environmental Conditions
Environmental Conditions
Replaceable Fuses
Equipment Class Class 1 Laboratory Equipment
Pollution Degree 2
Installation Category II
Environment Indoor Use
Altitude Not to exceed 3000 m
Electrical supply 200 - 240 V AC, 50/60 Hz, 2500 VA
Mains supply voltage Fluctuations not to exceed 10% of nominal supply
voltage
Operating Temperature 15 to 35°C (59 to 95°F)
Humidity < 85% RH at 35°C
T8A 250V 2110-0969
T12.5A 250V 2110-1398