Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 43
Maintenance 2
To check calibrant levels
To check calibrant levels
When required Monthly or weekly if you tune the instrument frequently
Tools required None
Parts required None
1 Examine each calibrant bottle. Enough tuning mix must be present to
immerse the end of the intake tube.
2 If the tuning mix level is within a few millimeters of the end of the intake
tube, refill the calibrant bottle.
NOTE
Record this procedure in the Maintenance Logbook.