138 Agilent 6500 Series Q-TOF LC/MS Maintenance Guide
2Maintenance
To reinstall the high pressure ion funnel process
To reinstall the high pressure ion funnel process
Parts Needed Needle nose pliers
Torx T-20 wrench
1 Install the ion funnel into the housing and secure with the two captive
screws.
2 Connect the five internal wires to the feed throughs:
Connect the two orange wires to the top connectors.
Connect the black wire to the bottom-rear connector.
Connect the red wire to the front-rear connector.
Connect the white wire to the bottom-most connector.
See Figure 75.
3 Install the desolvation assembly with two captive M4 screws.
4 Install the ESI with Agilent Jet Stream Technology.
5 Pump down the instrument.
Figure 75 Internal wires to the feed through.
Rear
Front
Orange Orange
White
RedBlack