EasyManuals Logo

Agilent Technologies 6500 Series User Manual

Agilent Technologies 6500 Series
154 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #49 background imageLoading...
Page #49 background image
Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 49
Maintenance 2
Electrospray Ion Source
Electrospray Ion Source
This section describes the removal, disassembly, cleaning, and reassembly of
the API-interface and the assemblies that make up the source.
To flush the nebulizer
When required Daily or at the end of each shift (or more often) to flush traces of samples and
buffers out of the tubing, valves and nebulizer.
Tools required Acetonitrile, HPLC-grade or better
Water, HPLC-grade or better
Parts required None
1 Make sure acetonitrile and water are two of the solvents installed in your
liquid chromatograph.
2 Set the liquid chromatograph to pump a mixture of 90% acetonitrile and
10% water at 2 ml/minute.
3 Pump this mixture through the nebulizer for 3 minutes.
NOTE
This procedure applies to both electrospray and APCI nebulizer.
NOTE
This is a good general-purpose flushing mixture but you may need to adjust it based on the
solvents, samples and buffers you are using. For example, a mixture of 50% acetonitrile and
50% water works well for removing salts.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6500 Series and is the answer not in the manual?

Agilent Technologies 6500 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6500 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals