144 Agilent 6500 Series Q-TOF LC/MS Maintenance Guide
2Maintenance
To replace the rough pump fluid
To replace the rough pump fluid
Replace the pump fluid every six months. Replace it sooner if the fluid
appears dark or cloudy.
Before you begin, make sure you have:
Container for catching old pump fluid
Funnel
Gloves, chemical resistant, clean, lint free (p/n 9300-1751)
1 Turn off the instrument.
See “To shut down the instrument” on page 19.
2 Unplug the power cord(s) from the instrument
Leave the power cord(s) unplugged while performing this procedure.
WARNING
W
Wear chemical-resistant gloves and safety glasses (goggles) for your safety.
WARNING
Never add or replace the rough pump fluid while the pump is on. Hot oil can splash
out and cause harm.
WARNING
The fill cap and pump may be dangerously hot. Check that the fill cup and pump are
cool before you touch them.
WARNING
Do not touch the fluid. The residues from some samples are toxic. Properly dispose
of the fluid.
CAUTION
Use only the rough pump fluid appropriate for your pump (Inland 45 oil for Edwards
E2M28). Any other fluids can substantially reduce pump life and invalidates the pump
warranty.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6500 Series and is the answer not in the manual?

Agilent Technologies 6500 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6500 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals