EasyManuals Logo

Agilent Technologies 6500 Series User Manual

Agilent Technologies 6500 Series
154 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #95 background imageLoading...
Page #95 background image
Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 95
Maintenance 2
To adjust the APCI nebulizer needle
To adjust the APCI nebulizer needle
When required After replacing the APCI nebulizer needle or if symptoms indicate the needle
may not be correctly adjusted.
Tools required Adjustment fixture (p/n G1946-20215)
Gloves, clean
Magnifier (p/n G1946-80049)
Wrench 3-mm, open-end (p/n 8710-2699)
Wrench ¼-inch x 5/16-inch, 2 required (p/n 8710-0510)
Parts required None
1 Install the nebulizer in the adjustment fixture.
Figure 40 Installing the nebulizer in the adjustment fixture
2 Loosen the needle holder locknut.
3 Position the magnifier so you can view the tip of the nebulizer.
4 Adjust the needle holder until the needle is even with the tip of the
nebulizer.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6500 Series and is the answer not in the manual?

Agilent Technologies 6500 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6500 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals