Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 29
Basic Operation 1
To install the Agilent Jet Stream inlet assembly
To install the Agilent Jet Stream inlet assembly
This topic applies to both the single and the dual Agilent Jet Stream source.
1 Install the Agilent Jet Stream spray shield.
2 Attach the Agilent Jet Stream inlet assembly to the desolvation chamber.
3 Close the latch on the Agilent Jet Stream inlet assembly.
4 Remove the plug from the reference nebulizer gas line and connect the
sheath gas tubing to the gas line.
5 Connect the electrical connections.
6 Connect the nebulizer gas line and LC tubing to the nebulizer.
7 On the Q-TOF tab, on the left side of the Method Editor window, select AJS
ESI or Dual AJS ESI, depending on your source.
8 Click Apply. The sheath gas turns on.
9 Click On to turn on the instrument. The Agilent Jet Stream inlet assembly
will begin to heat to the setpoints.
10 Load an appropriate method or create one, and allow the system to
stabilize.
11 If needed, install the ESI-L tuning mix in the reference B port of the CDS.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6500 Series and is the answer not in the manual?

Agilent Technologies 6500 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6500 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals