102 Agilent 6500 Series Q-TOF LC/MS Maintenance Guide
2Maintenance
Multimode Source and HPLC-Chip/MS Interface
Multimode Source and HPLC-Chip/MS Interface
This section describes how to clean and adjust the multimode source and
maintain the HPLC-Chip/MS interface.
To clean the multimode source daily
When required Daily or anytime you suspect carryover contamination from one sample or
analysis to another, or when you must access the end cap and capillary cap for
cleaning and inspection.
Tools required Gloves
Wash bottle, clean
Parts required Abrasive paper, 4000 grit (p/n 8660-0827)
Cloths, clean, lint-free (p/n 05980-60051)
Cotton swabs (p/n 5080-5400)
Mobile phase from the current method or clean isopropanol, reagent grade
or better
Water, reagent-grade or better
1 Turn off the spray chamber, nebulizer pressure, drying gas flow, drying gas
temp, and vaporizer temp.
2 Remove the nebulizer and the APCI corona needle.
3 Remove the cosmetic cover. You will have to remove the thermocouple
probe before you can wipe the spray chamber. Otherwise, leave the
thermocouple intact.
4 Open the spray chamber.
WARNING
Do not touch the multimode source or the capillary cap. They may be very hot. Allow
the multimode source to cool down before you handle them.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6500 Series and is the answer not in the manual?

Agilent Technologies 6500 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6500 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals