Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 25
Basic Operation 1
To remove a dual nano ESI source
To remove a dual nano ESI source
1 Put instrument in Standby mode.
2 Remove the reference and LC nanospray needle assembly from the source.
3 Clean both nanospray needles with 100% acetonitrile. You can also use any
solvent that dissolves the last sample run.
4 Disconnect the power cable from the lamp.
5 Open the spray chamber.
6 Remove the spray chamber by lifting it along the axis of its hinges.
7 Remove the capillary cap and drying gas diverter.
8 Install the standard capillary cap and spray shield. For the standard spray
shield, the small drying gas exit hole should be oriented above the main
hole, or entrance to the capillary cap.
WARNING
Do not touch the top of the dual nanospray source or the capillary cap. They may be
very hot. Let the parts cool before you handle them.
CAUTION
Do not let the sample recrystallize inside the needle, which leads to plugged needles.
Follow the steps above to prevent plugged needles.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6500 Series and is the answer not in the manual?

Agilent Technologies 6500 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6500 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals