Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 63
Maintenance 2
To reinstall the electrospray nebulizer
To reinstall the electrospray nebulizer
When required As needed.
Tools required None
Parts required None
1 Insert the nebulizer part way into the spray chamber.
2 Reconnect the nebulizing gas tubing to the nebulizer.
3 Finish inserting the nebulizer into the spray chamber.
4 Turn the nebulizer clockwise and lock it in place.
5 Reconnect the LC tubing to the nebulizer.
6 Close the nebulizer cover.
Do not hit the tip of the needle as you insert the nebulizer. The tip of the needle is easily
damaged.
Do not overtighten the LC fitting. Overtightening the fitting can crush the tubing,
creating a restriction.