32 Agilent 6500 Series Q-TOF LC/MS Maintenance Guide
2Maintenance
Calibrant Delivery System and Divert Valve
Calibrant Delivery System and Divert Valve
This section describes maintenance tasks that are related to the calibrant 
delivery system and divert valve.
When you do maintenance procedures on the divert valve:
• Do not bend any capillary line.
• The position of the divert valve can be set only in the On mode.
• Put the instrument in the Off mode.
• Remove the covers before you begin.
• Install the covers of the instrument after you finish.
Table 1 Divert Valve Connection (see Figure 2 and Figure 3)
No. Port Connection  User action
1 Inlet  from the front inlet union 
behind the inlet cover
none
2 Inlet from CDS none (permanent connection from calibrant delivery 
system)
3 Outlet  to the ion source connect the flexible capillary to the ion source
4 Loop loop to port 6 none (permanent connection) 
5 Waste  to the drain bottle  none
6 Loop loop to port 4 none (permanent connected)