Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 135
Maintenance 2
To clean the high pressure ion funnel
To clean the high pressure ion funnel
Parts Needed 2000 mL clean beaker
Isopropanol (IPA)
One pair of glove
1 Put the high pressure ion funnel on the table and on top of a clean cloth.
See Figure 72.
WARNING
This cleaning procedure requires the use of a sonicator and isopropanol. You must
clean the high pressure ion funnel in an approved fume hood with proper airflow
(lift) ), or you can experience a flash fire or explosion. The front window sash must
be in the closed position (down). All other electrical heating appliances such as hot
plates must be turned off.
WARNING
Wear chemical-resistant gloves and safety glasses (goggles) for your safety.