86 Agilent 6500 Series Q-TOF LC/MS Maintenance Guide
2Maintenance
APCI Source
APCI Source
This section describes how to open and close the APCI source and maintain it.
To learn how to install and remove the APCI source, see “To install an APCI
source” on page 27 and “To remove an APCI or APPI source” on page 23.
To clean the APCI spray chamber daily
When required Daily at the end of each shift or anytime you suspect carryover contamination
from one sample or analysis to another.
Tools required • Cloths, clean, lint-free (p/n 05980-60051)
• Gloves, clean
• Isopropanol, reagent grade or better
• Mobile phase, current
• Wash bottle, clean
• Water, reagent-grade or better
Parts required None
The APCI spray chamber operates at high temperatures. Allow sufficient time to
cool down before handling.
Some mobile phases are dangerous. Use the degree of caution appropriate for the
mobile phase being used.
Do not spray the mobile phase upward into the vaporizer.
Do not spray directly toward the tip of the capillary. This can cause pressure surges in
the vacuum system.