28 Agilent 6500 Series Q-TOF LC/MS Maintenance Guide
1 Basic Operation
To install an APPI source
To install an APPI source
1 Install the APPI USB to Serial Converter Cable (p/n 8121-1013) to the one
of the available USB ports on the SmartCard 4 Plus on the left side of the
Q-TOF instrument.
2 Rinse a clean bottle with acetonitrile.
The ship kit comes with an extra bottle (p/n 9300-2576) and bottle cap (p/n
9300-2575).
3 Pour the APCI/APPI Tuning Mix into the tuning mix bottle.
4 Attach the tuning mix bottle back onto the CDS.
5 Remove the currently installed source and install the G1971B APPI source
onto the spray chamber mounts.
6 Connect the APPI Vaporizor heater cable to its connector on the left side of
the Q-TOF. Then connect the APPI serial cable to the USB/Serial Converter
cable that was installed in step 1 of this procedure.
To install a nano ESI, dual nano ESI or multimode source or
HPLC-Chip/MS interface
Refer to the User Guide for the specific source or interface.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6500 Series and is the answer not in the manual?

Agilent Technologies 6500 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6500 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals