Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 35
Maintenance 2
To prepare tuning mix
To prepare tuning mix
For many instruments, the standard tuning mix will produce correct results.
However, in some instances, a dilution is required.
Dilute the tuning mix if:
You get error messages that instruct you to dilute the calibrant.
The tune report shows abundances greater than 750,000 counts. You
need to dilute until the largest abundance is below 750,000 counts for a
mass axis calibration. Typical tune mass abundances are in the range of
50,000 to 650,000 counts.
Do not dilute the tuning mix if:
Your source is an APPI or APCI.
Before you begin, make sure you have:
G1969-85000 ESI-L Low Concentration Tuning Mix (for Dual ESI or ESI
with Jet Stream Technology)
G1969-85020 MMI-L Low Concentration Tuning Mix (for multimode
source)
G1969-85003 Biopolymer Reference Mass Kit, which contains the
HP-0321 solution (for ESI with Jet Stream Technology)
HPLC-grade acetonitrile. Acetonitrile must be suitable for HPLC,
spectrophotometry, and pesticide residue analysis.
Water, 18 MΩ-cm resistivity or better and free of organics.
Glass CDS bottle (p/n 9300-2576) and cap (p/n 9300-2575)
The materials shipped with the instrument are of sufficient purity for this
purpose.
These instructions will make 100 mL of diluted tuning mix.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6500 Series and is the answer not in the manual?

Agilent Technologies 6500 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6500 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals