EasyManuals Logo

Agilent Technologies 6500 Series User Manual

Agilent Technologies 6500 Series
154 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #132 background imageLoading...
Page #132 background image
132 Agilent 6500 Series Q-TOF LC/MS Maintenance Guide
2Maintenance
High Pressure Ion Funnel (6550 only)
High Pressure Ion Funnel (6550 only)
With time and with heavy usage, the high pressure ion funnel needs to be
cleaned to maintain its performance on the 6550 Q-TOF.
To remove the high pressure ion funnel
Parts Needed Needle nose pliers
Torx T-20 wrench
1 Vent the LC/MS and unplug the system.
2 Let the ESI source with Agilent Jet Stream Technology cool down, then
remove it.
3 Remove the two M4 screws that hold down the desolvation assembly, then
remove the assembly.
4 Disconnect the five internal wires from the feed-throughs. Use needle nose
pliers. See Figure 70.
Connectors are fragile. Please be gentle when you remove these wires.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6500 Series and is the answer not in the manual?

Agilent Technologies 6500 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6500 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals