Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 97
Maintenance 2
To reinstall the APCI nebulizer
To reinstall the APCI nebulizer
When required As needed.
Tools required None
Parts required None
1 Insert the nebulizer into the spray chamber.
2 Turn it clockwise to lock it into place.
3 Reconnect the nebulizing gas tubing to the nebulizer.
4 Reconnect the LC tubing to the zero-dead-volume union.
5 Do not overtighten the LC fitting. Overtightening the fitting can crush the
tubing, creating a restriction.