Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 123
Maintenance 2
To clean the ion optics assembly
To clean the ion optics assembly
When required When the system has difficulty tuning, or when the system has poor
sensitivity.
Tools required Gloves, clean
Cloths, clean, lint-free (p/n 05980-60051)
Beakers, 500 ml, 2 ea
Tweezers
Screwdriver, TORX, T-10 (p/n 8710-1623)
Ball driver, 1.5 mm (p/n 8710-1570)
Methanol, reagent grade or better
Isopropanol, reagent grade or better
Acetone, reagent grade or better
Parts required None
1 Vent the system.
2 Remove the source.
3 Remove the desolvation assembly.
4 Remove the vacuum manifold cover.
5 Unplug all of the ion optics cables so that the ion optics assembly can be
removed. Pay close attention to the orientation of the cables and their
respective locations.
6 Use your finger to push on the skimmer spacer to get the ion optics to pop
out of the instrument. Be careful to catch it so it doesn't fall on the floor.
See Figure 59.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6500 Series and is the answer not in the manual?

Agilent Technologies 6500 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6500 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals