64 Agilent 6500 Series Q-TOF LC/MS Maintenance Guide
2Maintenance
To remove the desolvation assembly
To remove the desolvation assembly
When required When you need to access the optics assembly.
Tools required Wrench, ½-inch x 9/16-inch, open-end (p/n 8710-0877)
Screwdriver, TORX, T-20 (p/n 8710-1615)
Parts required None
1 Vent the system.
2 Unplug the instrument power cord from the power outlet after venting is
complete.
3 Remove the front, top and left side covers.
4 Disconnect the connections to the Aux module on top of the vacuum
manifold, and lift off and remove the Aux module. Disconnect the drying
gas tubing from the desolvation assembly.
5 Disconnect the drying gas heater cable from the desolvation assembly.
6 Disconnect the drain hose from the desolvation assembly.
7 Disconnect the PEEK nut on the tubing from the calibrant delivery system
to the desolvation assembly.
8 Remove the two retaining screws that keep the desolvation assembly
attached to the support rods.
WARNING
The spray chamber and desolvation assembly operate at very high temperatures.
Give them time to cool before proceeding.