Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 51
Maintenance 2
To clean the electrospray spray chamber daily
5 Rinse the interior of the spray chamber with the current mobile phase or
with a mixture of isopropanol and water.
6 Wipe the interior of the spray chamber with a clean, lint-free cloth.
7 Rinse the area around the spray shield.
8 Dampen a clean cloth with the mobile phase.
9 Wipe the spray shield and the area around the spray shield.
10 Close the spray chamber.
Figure 6 Opened electrospray spray chamber
WARNING
Some mobile phases are dangerous. Use the degree of caution appropriate for the
mobile phase being used.
CAUTION
Do not spray directly toward the tip of the capillary. This can cause pressure surges in
the vacuum system.