Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 137
Maintenance 2
To clean the high pressure ion funnel
3 Slowly pour water into the beaker until the solvent covers the ion funnel
completely. Sonicate the ion funnel for 5 minutes to remove any built up
salts or water soluble residue.
4 Pour the water out of the beaker and into an appropriate waste container.
5 Slowly pour isopropanol into the beaker until the solvent covers the Ion
Funnel completely.
6 Pour the isopropanol out of the beaker and into an appropriate waste
container.
7 Sonicate the assembly in a beaker of isopropanol for 15 minutes.
8 Remove the high pressure ion funnel from beaker and allow the isopropanol
to drain off the entire assembly.
9 Slowly pour methanol into the beaker until the solvent covers the ion funnel
completely. Sonicate the ion funnel for 5 minutes.
10 Use clean compressed nitrogen to blow out the remaining methanol from
between the funnel plates.
Figure 74 Pouring isopropanol into the glass beaker.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6500 Series and is the answer not in the manual?

Agilent Technologies 6500 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6500 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals