Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 105
Maintenance 2
To clean the multimode source weekly
6 Remove the spray shield. Use abrasive paper to gently clean the end of the
capillary cap.
7 Dampen a clean cloth and wipe the end of the capillary cap.
8 Reinstall the spray chamber.
9 Use abrasive paper to gently clean the spray shield. Dampen a clean cloth
and wipe the spray shield.
10 Rinse the area around the spray shield then wipe the area around the spray
shield.
11 Reinstall the spray chamber on the instrument.
12 Replace the nebulizer and APCI corona needle.
13 Install the thermocouple probe and adjust it so that it protrudes 15mm
from the inner spray chamber wall.
14 Replace the cosmetic cover.
15 Close the spray chamber.