Working with Application Graphs Applications
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave
Multichannel System, Fourth Edition
155
Working with Application
Graphs
If you press the [Graph] softkey, a graph similar to Figure 108 appears. This
section explains how to use the graph.
There are four markers A, B, C, and D. You can use these markers to choose the
area of the graph that you wish to zoom into.
For the Logging and Stability applications, the time values of A and B are
displayed at the bottom of the screen in units of seconds. The symbol that is
listed directly beneath these values, displays the difference between A and B in
seconds.
For the PACT application, the wavelength values of A and B are displayed at the
bottom of the screen in units of nanometers (nm). The symbol that is listed
directly beneath these values, displays the difference between A and B in
nanometers (nm).
The power values of C and D are displayed at the bottom of the screen in the
chosen power units. The symbol that is listed directly beneath these values,
displays the difference between C and D in the chosen power units.
NOTE As a default, the C and D markers are out of range when you first open a graph.
See How to Set Markers on page 156, for more information on setting moving
markers.
Figure 108 The Application Graph