Specifications Declaration of Conformity
242 Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement
System, & Agilent 8166A Lightwave Multichannel System, Fourth Edition
Agilent 8166A Lightwave Multichannel
System
We declare that the product:
conforms to the following standards:
Manufacturer: Agilent Technologies Deutschland GmbH
Optical Communication Measurement
Herrenberger Strasse 110-140
D-71034 Böblingen
Germany
Lightwave Multichannel System
consisting of: Agilent 8166A Mainframe
and modules: Agilent 81611A
1
1300 nm Return Loss Module
Agilent 81612A
1
1550 nm Return Loss Module
Agilent 81613A
1
1310/1550 nm Return Loss Module
Agilent 81614A
1
1550/1625 nm Return Loss Module
Agilent 81650A 1310 nm Laser Source Module
Agilent 81651A 1550 nm Laser Source Module
Agilent 81652A 1550/1625 nm Laser Source Module
Agilent 81653A 1650 nm Laser Source Module
Agilent 81654A 1310/1550 nm Laser Source Module
Agilent 81655A 1310 nm Laser Source Module
Agilent 81656A 1550 nm Laser Source Module
Agilent 81657A 1310/1550 nm Laser Source Module
Agilent 81689A
2
Compact Tunable Laser for Multichannel Test Applications
Safety: IEC 1010-1:1990 incl. A1:1992 + A2:1995 EN 61010-1:(1993) + A2:1995
IEC 825-1 (1993); FDA 21 CFR 1040.10 EN 60825-1 (1994)
EMC: radiation
EN 55011:1991 / CISPR 11:1990 mod.
Group 1, Class B (1)
EMC: immunity
EN 50082-1:1997
EN 61000-4-2:1995 ESD: 4KV cd, 8KV ad, 4KV cp
EN 61000-4-3: 1995 Radiated Immunity: 3 V/m 80%
ENV 50204: 1995 Radiated Immunity: 3 V/m 50% Duty
EN 61000-4-4:1995 Fast Transients/Bursts ±1 kV
EN 61000-4-5:1995 Surges ±1 kV; ±2 kV
EN 61000-4-6:1995 Conducted Immunity 3 V, 80%
EN 61000-4-8:1993 Power freq. Magn. Field 3A/m; 50 Hz
IEC 1000-4-11:1994
Voltage Dips and
Interruptions
30%-10 ms; 60%-100 ms