Cleaning Instrument Housings Cleaning Instructions
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave 
Multichannel System, Fourth Edition
273
Cleaning Instrument Housings
Use a dry and very soft cotton tissue to clean the instrument housing and the 
keypad. Do not open the instruments as there is a danger of electric shock, or 
electrostatic discharge. Opening the instrument can cause damage to sensitive 
components, and in addition your warranty will be voided.
Which Cleaning Procedure 
should I use ?
Light dirt If you just want to clean away light dirt, observe the following procedure for all 
devices:
• Use compressed air to blow away large particles.
• Clean the device with a dry cotton swab.
• Use compressed air to blow away any remaining filament left by the swab.
Heavy dirt If the above procedure is not enough to clean your instrument, follow one of the 
procedures below. Please consult “Cleaning Instructions for this Instrument” on 
page 266 for the procedure relevant for this instrument.
If you are unsure of how sensitive your device is to cleaning, please contact the 
manufacturer or your sales distributor
How to clean connectors
Cleaning connectors is difficult as the core diameter of a single-mode fiber is only 
about 9 µm. This generally means you cannot see streaks or scratches on the 
surface. To be certain of the condition of the surface of your connector and to 
check it after cleaning, you need a microscope.
In the case of scratches, or of dust that has been burnt onto the surface of the 
connector, you may have no option but to polish the connector. This depends on 
the degree of dirtiness, or the depth of the scratches. This is a difficult procedure