Tunable Lasers How to Perform a Wavelength Sweep
118 Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement 
System, & Agilent 8166A Lightwave Multichannel System, Fourth Edition
a when the sweep is completed, or 
b when you press [Stop].
How to Perform a Manual Sweep
You can perform a manual sweep if you choose the <Manual> sweep mode or if 
you press [Pause] during an automatic sweep. In a manual sweep you choose 
when you want to perform each step of a stepped sweep. You can choose to move 
forward or backward a wavelength or to end the sweep. 
To perform a manual wavelength sweep over the range 1510 nm to 1570 nm, 
three times, sweeping two ways, in 1 nm steps:
1 Move to the Tunable Laser channel and press [Details].
2 Move to [Sweep Mode] and press Enter.
3 Move to the <Manual>, by using the cursor key, and press Enter.
4 Perform step 4 to step 13 on page 114 to set the sweep parameters.
5 Press the [Active] hardkey on the front panel of your Tunable Laser module to 
enable the optical output.
6 Press [Run Swp] to start the sweep. The screen in Figure 76 then appears.
7 You can:
– press [Next] to move on to the next wavelength step,
– press [Prev] to move on to the previous wavelength step, or
– press [Stop] to end the sweep.