EasyManuals Logo

Agilent Technologies 8163A User Guide

Agilent Technologies 8163A
295 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #151 background imageLoading...
Page #151 background image
A Background to Return Loss Measurement with the HP 81534A Return Loss Measurement
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave
Multichannel System, Fourth Edition
151
Calculating the Insertion Loss of the DUT
Insertion Loss is explained in What is Insertion Loss ? on page 134.
To measure the insertion loss you must measure the power transmitted through
the measurement cable, see Measuring the Power Transmitted Through the
Measurement Cable on page 146 and the power transmitted through the DUT,
see Measuring the Power Transmitted Through the DUT on page 147.
The equation below gives the insertion loss of the DUT, IL
DUT
:
IL
DUT
10
E
il
E
Meas
---------------log=
(18)

Table of Contents

Other manuals for Agilent Technologies 8163A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8163A and is the answer not in the manual?

Agilent Technologies 8163A Specifications

General IconGeneral
BrandAgilent Technologies
Model8163A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals