Working with Application Graphs Applications
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave
Multichannel System, Fourth Edition
161
How to Lock Markers to the Curve
You can lock markers A and B to curves by performing the following
instructions:
1 Press the [Menu] softkey and move to <Lock to Curve>. A menu displaying
the options displayed in The Lock to Curve Menu on page 161 appears.
2 Press the [A] softkey to lock marker A to a curve. The number opposite
Marker A on curve:
is highlighted.
3 Use the cursor key to move to the legend number of the curve to which you
want to lock marker A.
4 Press Enter. Marker B will be locked to the same curve as marker A, if it was
assigned to curve 0 immediately before, that is, it was unassigned immediately
before.
5 If you want to assign marker B to another curve, press the [B] softkey, use the
cursor key to move to the legend number of the curve to which you want to
lock marker B, and press Enter.
6 Press Close to return to the graph.
How to Unlock Markers from the Curve
You can unlock markers A and B from the curve by performing the following
instructions:
1 Press the [Menu] softkey.
2 Move to <Unlock From Curve> and press Enter to return to the graph.
Figure 113 The Lock to Curve Menu