How to Use Laser Source Modules Laser Sources
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave
Multichannel System, Fourth Edition
95
How to Set Attenuation
The amount of power that is output from a source can be controlled. You can use
the attenuation parameter to reduce the optical power of the laser output.
P
output
= P
source
− ATT (dB)
Where,
P
output
is the power level at the output of the module,
P
source
is the power level at the output of the source, and
ATT is the attenuation parameter.
To set the attenuation value to 2 dB:
1 Move to the laser source channel and press the [Details] softkey. You see the
screen in Figure 62.
2 Move to the [Attenuation] parameter and press Enter.
3 Enter 2.00 and press Enter.
How to Modulate the Optical Output
How to Change Modulation Source
The Modulation Source parameter, [Mod Src], allows you to modulate the output
signal.
Figure 62 The Power Sensor Details Screen

Table of Contents

Other manuals for Agilent Technologies 8163A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8163A and is the answer not in the manual?

Agilent Technologies 8163A Specifications

General IconGeneral
BrandAgilent Technologies
Model8163A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals