Applications The Stability Application
174 Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement 
System, & Agilent 8166A Lightwave Multichannel System, Fourth Edition
The Stability application differs from the “The Logging Application” on 
page 164 because you set the averaging time, the total time, and the maximum 
number of measurements and all power measurements and the instrument 
calculates a period time, a new measurement is started after the completion of 
every period time. Period time may be greater than or equal to the averaging time 
you set and is always greater than or equal to 100 ms. 
This leaves the following two possibilities:
• The averaging time is less than the period time, there is a dwell time between 
the commencement of each power measurement, as shown in Figure 126. This 
can happen because:
– the maximum number of power measurements multiplied by the averaging 
time is less than the total time or 
– the averaging time is less than 100 ms.
• The averaging time is equal to the period time, all power measuremant are 
performed without any dwell time. 
NOTE The total time concludes after the last dwell time period, not after the last 
averaging time period.
The Stability application enables you to make periodic power measurements over 
a relatively long time period (23 hours, 59 minutes and, 59 seconds).
The Stability application differs from “The Logging Application” on page 164 
because you may use auto-ranging mode, see step 9 on page 177 for more 
information.
All results can be: 
• displayed on the screen,
• printed out to hardcopy, or
• saved to disk drive of your Agilent 8164A Lightwave Measurement System.
How to Set Up a Stability Function
To set up a stability function:
1 Press the Appl hardkey. The Applications menu, as shown in Figure 117, 
appears.