EasyManua.ls Logo

Agilent Technologies 8163A - Setting Logging Parameter; Running a Logging Application

Agilent Technologies 8163A
295 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Applications The Logging Application
168 Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement
System, & Agilent 8166A Lightwave Multichannel System, Fourth Edition
12 If you have chosen <common> as the <Range Mode>, move to a range setting
and press Enter.
If you have chosen <individual> as the <Range Mode>, a screen displaying all
the selected Power Meter channels and their corresponding range settings is
displayed. For each selected Power Meter channel, press Enter, move to a
range setting, and press Enter.
NOTE If you have chosen <individual> as the <Range Mode> and want to access the
screen displaying all the selected Power Meter channels and their
corresponding range settings from the Logging Setup screen, move to [Range],
press the [Edit] softkey.
13 Move to <Ref Mode>, press Enter, move to one of the following referencing
modes:
<Val u e>, you set a reference value using <Ref>, see step 14,
<First Sample>, the first sample for each channel of the logging function is
selected as the reference value for all subsequent measurements for that
channel,
<Channel 1>, the value that Channel 1 measures is updated as the reference
value for each set of power measurements (thus, each data value for
channel 1 will be 0 dB), and
press Enter.
NOTE These settings only have an effect when you choose <dB> as the <Pwr unit>.
See step 8 or step 22.
14 If you have chosen <Valu e> as the <Ref Mode>, move to <Ref>, press the
[Edit] softkey, enter a value for the reference, and press Enter.
Setting Logging Parameter
15 Move to [Samples], press Enter, enter an integer value between 1 and 4000,
and press Enter.
16 Move to <AvgTime>, press Enter, move to an averaging time period, and press
Enter.
Running a Logging Application
17 You can run the logging application by pressing [Measure]. The Logging
Measurement screen, as shown in Figure 121, appears. A progress bar at the
bottom of the screen shows the proportion of the logging application that has
been completed.

Table of Contents

Other manuals for Agilent Technologies 8163A

Related product manuals