Tunable Lasers How to Use Triggers
126 Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement
System, & Agilent 8166A Lightwave Multichannel System, Fourth Edition
3 Move to <Input Trigger Mode>, by using the cursor key, and press ENTER.
You will see the screen in Figure 81.
4 Move to one of the following, by using the cursor key:
<Ignore>, input triggers are ignored.
<Next Step>, an input trigger will cause the next step of a stepped sweep to be
performed.
<Run Sweep>, an input trigger will start a single sweep cycle.
5 Press Enter.
You can generate input triggers in any of the following ways:
applying a trigger to the Input Trigger Connector on the rear panel of your
instrument,
setting <Trigger Configuration> to <Loopback> so that an output trigger
automatically generates an input trigger, or
using the
:TRIGger
GPIB command, see your mainframes Programming
Guide.
How to Use Output Triggering
You can configure your Tunable Laser module to output a trigger when the
instrument performs certain tasks.
To set your modules Output Trigger Configuration:
1 See How to Set the Trigger Configuration on page 62 for how to configure
the trigger connectors.
Figure 81 Input Trigger Mode

Table of Contents

Other manuals for Agilent Technologies 8163A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8163A and is the answer not in the manual?

Agilent Technologies 8163A Specifications

General IconGeneral
BrandAgilent Technologies
Model8163A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals