EasyManuals Logo

Agilent Technologies 8163A User Guide

Agilent Technologies 8163A
295 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #115 background imageLoading...
Page #115 background image
How to Perform a Wavelength Sweep Tunable Lasers
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave
Multichannel System, Fourth Edition
115
6 Move to [λ Stop] and press Enter.
7 Enter 1570.000, press Enter.
8 Move to [Cycles] and press Enter.
9 Enter 3 and press Enter.
10 Move to [Repeat Mode] and press Enter.
11 Move to <Twoway>, by using the cursor key, and press Enter.
12 Move to [Step] and press Enter.
13 Enter 1.000 and press Enter.
14 Move to [Dwell] and press Enter.
15 Enter 0.5000 and press Enter.
16 Press the [Active] hardkey on the front panel of your Tunable Laser module to
enable the optical output.
17 Press [Run Swp] to start the sweep. The screen in Figure 74 appears.
18 The wavelength is swept automatically but you can:
a press [Stop] to end the sweep, or
b press [Pause] to pause the sweep. The screen in Figure 75 appears. You
can:
– Press [Continue] to continue sweeping automatically.
– Press [Prev] or [Next] to sweep manually.
– Press [Stop] to end the sweep.

Table of Contents

Other manuals for Agilent Technologies 8163A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8163A and is the answer not in the manual?

Agilent Technologies 8163A Specifications

General IconGeneral
BrandAgilent Technologies
Model8163A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals