aurora
Sonata
Hand Held ISDN Tester—User Guide
4-26 427869
-1-
Select Threshold from the BERT Setup menu.
The BER display shows the error rate that applies if the
currently selected number of errors occur within the selected
test time. The rate is shown as x.yE-z, where E-z means 10 to
the power of z. In the above example, the rate is 2.6x10
7
.
You can set the threshold either in terms of a number of errors
or as an HRX percentage. HRX (Hypothetical Reference
Connection) is a model used to study the performance of
ISDN, where the percentages represent the allocation of
degraded minutes and errored seconds. For details, refer to
ITU-T Recommendation G.821.
-1-
Select Mode from the BERT Setup menu to
display an option window.
-2-
Choose either Bit Errors—set a number of
errors as the threshold, or HRX—choose an
HRX percentage, and press ENTER. The
option window closes.
To set the threshold as a number of errors (i.e. Mode is set to
Bit Errors):
-1-
Select Errors from the BERT Setup menu to
display an entry window.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies auroraSonata and is the answer not in the manual?

Agilent Technologies auroraSonata Specifications

General IconGeneral
BrandAgilent Technologies
ModelauroraSonata
CategoryTest Equipment
LanguageEnglish

Related product manuals